Basic operations – RIGOL DS6000 Series User Manual

Page 33

background image

RIGOL

RP1000D User’s Guide

5

Basic Operations


1. Connect the red safety IC clip with one end of the red

dual-banana plug silicon cable and the black safety IC clip with
one end of the black dual-banana plug silicon cable provided in
the accessories. Then, connect the red dual-banana plug silicon
cable with the red (+) input terminal of the high voltage probe
and the black dual-banana plug silicon cable with the black (-)
input terminal of the high voltage probe.

Note:
a) The safety IC clip can be replaced by the high voltage

dedicated IC clip, safety alligator clip or safety contact
probe prod;

b) The dual-banana plug silicon cable can be replaced by high

voltage dedicated dual-banana plug silicon cable.


2. Connect one end of the dual-BNC coaxial cable to the BNC

interface of the high voltage differential probe and the other
end to input terminal of the oscilloscope.

3. Turn on the channel switch on the oscilloscope and adjust the

high voltage probe and oscilloscope to make the attenuation
ratios of the two match. If the attenuation ratio of the
oscilloscope does not match that of the high voltage probe, the
actual vertical scale equals the attenuation ratio of the high
voltage probe divided by the attenuation ratio of the
oscilloscope and then times the vertical scale of the
oscilloscope.