beautypg.com

Get the results you need, Field check gauge weekly – Retrotec DucTester 351 User Manual

Page 4

background image

Flow at the induced
pressure is the simplest
result.

Flow in cubic meters
per hour per square
meter (m

3

/(h·m

2

) may

be required.

(Enter an area)

Flow in cubic meters
per second per square
meter (m

3

/(s·m

2

) may

also be required.

(Enter an area)

Flow reads “--” at test pressure?
If the test pressure has been reached, but “--” appears,

the fan is running too slowly to accurately measure flow.

Switch to a lower Range Plug

Change [Range] to match

Re-adjust fan speed.

Cannot achieve test pressure at full speed?
If fan reaches 100% speed before reaching 80 Pa:

Change to a higher Range Plug and try again.

Change [Range] on the gauge to match.

Check seals on all registers. Look for disconnected

ducts or ducts open to outdoors.

Tap [@ Pressure] to get the gauge to calculate

what the flow would be at exactly 80 Pa.

Page 4 of 4

Get the results you need

Tap the [Channel B] key to select a different

Result, or tap the [Result to be displayed] key

on the [Settings] menu.

Show leakage area Result

Step 5:

Desired results not achieved?

Equivalent Leakage Area (EqLA) describes the leakage

area in terms of one large hole in a flat surface.
Tap the [Channel B] key, then

, and select

“EqLA: cm”

[Channel A] shows the duct pressure and

[Channel B] “EqLA” shows the combined size of all

the holes in the ducts.

Leakage area is not a required result, but is a nice way to
visualize the size of the hole in the ducts.

When a Result is chosen that requires an area or volume,

or

will be shown on the Home screen. Tap to change.

The area and volume can also be changed from the

menu.

Check gauge operation and check for blocked, leaking or

pinched tubes weekly, and anytime results are in question.

To perform the gauge check, you will need the gauge

and Umbilical.

Set [Time Average] to 5 seconds in [Settings].

Tap [Channel B] and select“Pressure: Pa”.

Connect the yellow tube between the red and

yellow ports.

Repeat between different ports with each of the

tubes you use for testing.

Field check gauge weekly

If readings on Channel A and Channel B are within

2% and don’t drop rapidly, the tube is not blocked

or leaking and the gauge is correct.

Checking your gauge

and tubes regularly will

eliminate a common

source of error in

readings.