Warning:

At least one fan must be enabled for

correct DM-2 behavior!

The Device Setup sub-menu allows fans and ranges

to be added or removed from the selections available

when [Device] is pressed. Eliminate all unused Devices

to avoid confusion.

● Press [Setup] and then [

]

until “Device Setup”

appears. Press [Enter] to view

available Devices.

● Scroll through the Devices.

Press [

] to select “Yes” to add, or “No” to remove.

● Press [Exit] twice to return to the main screen.

Enable:

●“Retrotec DU200” for DU200 DucTester system

●“Retrotec 1000” for 1000 systems

●“Retrotec 2000” for Q46 and Q56 systems

●“Retrotec 3000SR” for Q4E, Q5E, and QMG Door

Fan systems with green and yellow connection

ports on the fan.

Unused ranges may be removed to speed up cycling

between useful ranges.

● Select a fan from the Device Setup sub-menu,

press [Enter] to view the possible ranges for

that device.

● Scroll through the possible ranges. Press [

] to

select “Yes” to add a range, or “No” to remove a range.

● Press [Exit] once to return to the Device Setup

sub-menu, three times to return to the main screen.

Use the Mode Setup sub-menu to add or remove

displayed results (Flow, EqLA, ACH, etc.), to select units

and to change the default extrapolation pressures that

appear when [@ Pressure] is used.

● Press [Setup] and then [

] until “Mode Setup”

appears. Press [Enter] to view all modes that will

be available when [Mode] is pressed.

● Scroll through possible modes. Press [

] to select

units or select “Off”; that mode will not appear.

● Most modes have an associated “@Pressure”,

the pressure to which the displayed results will be

extrapolated when [@ Pressure] is pressed.

● To change the “@Pressure”, press [Enter] and input

a value between [1 ‒ 200], then press [Enter]

again to save.

● Press [Exit] twice to return to the main screen.

● Pressure cannot be disabled. All other results can

be disabled.

QuickGuide DM-2

The Setup Menu provides configuration

options for all of the DM-2’s functions

and capabilities. Depending on the

firm ware version, some options may not

be available. Properly setting up the

gauge can eliminate confusion and incorrect results.

● Press [Setup] to view the main setup menu.

● Use the [

], [

], to navigate around the menu. Press

[

], and [

] to change the current selection.

Press [Enter] to enter a sub-menu or to change a

configurable numeric value.

n: sets the slope of the @-Pressure

extrapolation line. Use 0.65 for tight houses

and 0.60 for ducts and large buildings.

Power Down Hour: If the DM-2 remains

inactive for this length of time, it turns off

automatically. The DM-2 will never turn off

if a value of “0” is entered.

Surface Area Unit: For building envelope or

floor areas used in Flow or Leakage

Area per Building Area results.

Building Volume Unit: Units for volume

used in air-change calculations.

European, Separator: “No” displays one

thousand as “1.000”, “Yes” displays one

thousand as “1,000”.

Sig Figs: Eg. 1134 is displayed as 1135

for 3.5 Significant Figures, 1130 for 3.0

and 1150 for 2.5.

Device Setup: Add or remove Devices

and Ranges.

Mode Setup: Add or remove Modes, set units.

Full Screen Timeout: If no keys are pressed,

the DM-2 will change to the large format

screen after this

length of time.

Restore Settings:

Select “Yes” to restore

the gauge to factory

settings upon leaving

the Setup Menu.

Language: Choose the language displayed.

Battery Type: Choose rechargeable or

not rechargeable.

Display Version Info: Select “Yes” to view

the current DM-2 version information. Press

[Exit] to return to the Setup Menu.

● Press [Exit] to return to the main screen.

Add/remove Devices

Add/remove Ranges

Enable/disable results Mode

Set up gauge