Quickguide dm-2, Step 8: discover autozero, Step 2: exit – Retrotec DM-2 Gauge User Manual

Page 2

background image

Tip: Leaving the backlight on discharges the battery at

a faster rate.

1s, 2s, 4s, 8s: Display updates every 1 second.

10s, 20s, 1min, 2min: Display updates every 10 seconds.

Tip:

Auto Zero can be turned off to conserve batteries

when not taking readings.

Tip: By default, Flow, Air Changes, and EqLA are

enabled. All other results can be enabled manually

from the Setup Menu. See Page 4.

Tip: From the factory, the Retrotec DU200, Retrotec

1000, Retrotec 2000 and Retrotec 3000SR are enabled.

All other Devices are disabled. See Page 4 to add or

remove Devices to save time and reduce errors.

Tip: Disable the other Flow Ranges using [Setup] to

save time and reduce errors.

● Press [Baseline] to begin

sampling and press [Enter]

after 15 to 60 seconds.

● The displayed Baseline pressure is subtracted

from all future readings.

● Pressing [Exit] at any time will clear the Baseline.

● Press [On/Off] to turn the DM-2 On.
● Press and hold [On/Off] to turn the

DM-2 Off.
● When gauge is on press [On/Off] to
turn the backlight on/off.

● Press [Exit] to:

● Clear the splash screen.

● Stop the fan and cancel Set Speed or

Set Pressure mode.

● Clear and cancel the set Baseline.

● Move back or exit the Setup menus.

● Cancel Volume and Area inputs.

● Cancel other data entry.

● Press [Enter] to acknowledge inputs

and allow data entry:

● Change volume in Air Chg mode.

● Change area in Flow/Area and

Leakage/Area modes.

● Accept indicated Baseline pressure.

● Accept input fan speed or pressure.

● Press [Mode] repeatedly to scroll

through the enabled results.

● The DM-2 supports on-the-fly

calculation of all currently used

results worldwide, including Flow, Air

Changes, Leakage Areas, Normalized Leakage Area,

Specific Leakage, Permeability and many others.

Results are averaged over the selected

time period

● Time averaging should always be On.

● Press [Time Avg] to scroll through the possible

time periods.

When AutoZero is active, the gauge

automatically zeros itself every 8

seconds, even while making

measurements.

● Press [Auto Zero] until “On” appears.

● AutoZero must be left on under normal operation.

Remove bias/offset/background/Baseline

pressures due to wind or stack that

are present before test pressures

are applied.

Step 3:

Enter

Step 6:

Select the Mode

Step 1:

On/Off/Backlight

Step 7:

Explore Time Averaging

Step 8:

Discover AutoZero

QuickGuide DM-2

The DM-2 calculates flows for all

Retrotec, TEC and Infiltec fans.

Choosing the right fan or flow device

is required to get correct results.

● Press [Device] to scroll through the enabled fans.

Step 4:

Select Device

Step 9:

Learn about Baseline mode

Step 5:

Select fan Range

● Press [Range Config] to scroll through

the enabled Flow Ranges for the

currently selected device.

Warning:

Selecting the incorrect range will yield

incorrect results.

Step 2:

Exit