5
CAUTION
DO NOT place any liquids on or near the Concentrator. If any liquid
gets on or in the Concentrator, immediately turn “OFF” the power
switch, unplug from the electrical outlet and contact your Provider .
SPECIFICATIONS
Dimensions:
Length:
17.1 in
(43.5 cm)
Width:
12.3 in
(31.1 cm)
Height:
17.0 in
(43.2 cm)
Weight:
29.6 lbs
(13.4 kg)
Shipping Weight:
33.9 lbs
(15.4 kg)
Altitude*:
Sea Level to 10,000 ft (3,048 m)
Operating Temperature:
50˚F to 95˚F
(10˚C to 35˚C)
Storage/ Transport Temperature: -29˚F to 140˚F (-34˚C to 60˚C)
Operating / Storage / Transport
Relative Humidity:
Up to 95% Non-condensing
Outlet Pressure:
8.0 psi ±10%
Oxygen Concentration and Flow Tolerance:
*Altitude
Flow Tolerance (l/min)
Oxygen %
Up to 6,500 ft
± 10% of settings 2 - 5
± 200 ml settings 0.5 -1.5
87 - 95 %
6,501 - 10,000 ft
± 20% of settings 1 - 5
± 200 ml settings 0.5
85 - 90 %
When checking the concentrator’s performance, the oxygen analyzer
used should comply with ISO 21647 standard.
Start up time:
87% O
2
Concentration
≤ 3 min
Electrical (AC Power): 115 VAC ±10%, 60 Hz, 3.2 Amps; 350 Watts
Tubing Limitations:
7 ft cannula with a maximum 40 ft of Crush-
Proof Tubing (47 ft total).
Equipment Classification:
Class II
- with respect to protection from electrical shock
Type BF
- degree of protection against electrical shock
IPX1
- degree of protection against ingress of liquids
Mode of Operation
- Continuous
This Concentrator meets the requirements of IEC 60601-1, IEC 60601-1-2,
IEC 60601-1-4 and Safety Requirements for ISO 8359.