DANGER

• DO NOT smoke, allow others to smoke or have open flames near the

Concentrator when in use.

• DO NOT use near any type of flame or flammable/explosive substances.

• DO NOT allow Concentrator to exceed “Operating or Storage

Temperature Specifications”.

• The Concentrator is not suitable for use in the presence of flammable

anesthetic.

TO REDUCE THE RISK OF ELECTRICAL SHOCK:

• DO NOT use while bathing.

• DO NOT reach for a Concentrator that has fallen into water.

Unplug immediately!

• DO NOT attempt to repair Concentrator. Any attempt could result in

electrical shock.

WARNING

• Use the Concentrator for its “Intended Use” as described in this manual.

• In the event of an equipment audible alarm or if you are experiencing any

signs of discomfort, connect to another source of oxygen and contact

your Provider and/or Health Care Professional immediately.

• In certain circumstances oxygen therapy can be hazardous; any user

should seek medical advice prior to using this Concentrator.

• Where the prescribing Health care professional has determined that an

interruption in the supply of oxygen, for any reason, may have serious

consequences to the user, an alternative source of oxygen should be

available for immediate use.

• ALWAYS keep the Concentrator in an upright position.

• ALWAYS keep at least 6 in. (15.24 cm) away from walls, furniture, and

especially curtains that could prevent adequate airflow to the Concentrator.

• DO NOT place the Concentrator in a small closed space such as a closet.

• DO NOT use extension cords or electrical adapters.

• The Concentrator is designed to operate with a single lumen, adult cannula.

• Flow is available ONLY at the listed settings.

There is NO FLOW between settings.

• Flow Setting Knob does not rotate 360 degrees.

TO REDUCE RISK OF BURNS, ELECTROCUTION, FIRE, OR INJURY

TO PERSONS:

• Keep the Concentrator away from heat or open flame.

• DO NOT use if Concentrator has a damaged power cord or plug.

• DO NOT use if Concentrator has been dropped or damaged.

• DO NOT use oil or grease on the Concentrator.

• DO NOT cover the air vents, the Concentrator requires proper ventilation.

• DO NOT remove the cover on the Concentrator. Servicing must be