Maintenance / cleaning, Replacement parts, Returns – Precision Medical PM4351 EasyFlow5 User Manual
Page 11: Disposal instructions, Warning

10
MAINTENANCE / CLEANING
WARNING
DO NOT remove cover or attempt to repair. There are no user serviceable parts
inside. Contact your Provider for repairs.
1. Patient should connect to an alternate Oxygen source.
2. Turn the Concentrator to the “OFF” ( ) position.
3. Unplug the Power Cord from electrical outlet.
4. Clean exterior surfaces of the Concentrator with a cloth dampened with water and
mild detergent.
5. Wipe the Concentrator dry with a clean cloth.
6. When not in use, store the Concentrator in a clean area free from grease, oil and
other sources of contamination.
Tell us how we are doing!
Visit us at www.precisionmedical.com
REPLACEMENT PARTS
DESCRIPTION
PART#
Humidifier Adapter
507324
RETURNS
Returned products require a Returned Goods Authorization (RGA) number, contact
Precision Medical, Inc. All returns must be packaged in sealed containers to prevent
damage. Precision Medical, Inc. will not be responsible for goods damaged in
transit. Refer to Precision Medical, Inc. Return Policy available on the Internet,
www.precisionmedical.com.
DISPOSAL INSTRUCTIONS
This device may contain electrical components that are hazardous to the environment.
DO NOT dispose device into standard trash. Contact your local waste Management for
disposal of Electronic Equipment.
Visual Indicator/
Problem
Audible
Alarm
Probable Cause
Remedy
Condensation
collects in the
oxygen tubing
when you use a
bubble humidifier
bottle.
None
Concentrator not properly
ventilated, elevated
operating temperatures.
Ensure Concentrator is
positioned away from curtains
or drapes, hot air registers or
heaters. Be certain to place the
Concentrator so all sides are at
least 6 in. (15.24 cm) away from
a wall or other obstruction. DO
NOT place the Concentrator in a
confined area.
DO NOT OVERFILL bubble
humidifier. Allow the oxygen
tubing to dry out, or replace with
new tubing.