MAINTENANCE / CLEANING

WARNING

DO NOT remove cover or attempt to repair. There are no user serviceable parts

inside. Contact your Provider for repairs.

1. Patient should connect to an alternate Oxygen source.
2. Turn the Concentrator to the “OFF” ( ) position.
3. Unplug the Power Cord from electrical outlet.
4. Clean exterior surfaces of the Concentrator with a cloth dampened with water and

mild detergent.

5. Wipe the Concentrator dry with a clean cloth.
6. When not in use, store the Concentrator in a clean area free from grease, oil and

other sources of contamination.

Tell us how we are doing!

Visit us at www.precisionmedical.com

REPLACEMENT PARTS

DESCRIPTION

PART#

Humidifier Adapter

507324

RETURNS

Returned products require a Returned Goods Authorization (RGA) number, contact

Precision Medical, Inc. All returns must be packaged in sealed containers to prevent

damage. Precision Medical, Inc. will not be responsible for goods damaged in

transit. Refer to Precision Medical, Inc. Return Policy available on the Internet,

www.precisionmedical.com.

DISPOSAL INSTRUCTIONS

This device may contain electrical components that are hazardous to the environment.

DO NOT dispose device into standard trash. Contact your local waste Management for

disposal of Electronic Equipment.

Visual Indicator/

Problem

Audible

Alarm

Probable Cause

Remedy

Condensation

collects in the

oxygen tubing

when you use a

bubble humidifier

bottle.

None

Concentrator not properly

ventilated, elevated

operating temperatures.

Ensure Concentrator is

positioned away from curtains

or drapes, hot air registers or

heaters. Be certain to place the

Concentrator so all sides are at

least 6 in. (15.24 cm) away from

a wall or other obstruction. DO

NOT place the Concentrator in a

confined area.

DO NOT OVERFILL bubble

humidifier. Allow the oxygen

tubing to dry out, or replace with

new tubing.