Step 2: cleaning, High flow – Precision Medical HeliO2 Blenders User Manual
Page 20

Heliox / Oxygen Blender Service Manual
Page 18
HIGH FLOW
Figure F
16. Remove each diaphragm housing assembly [20] from the manifold by
removing the two hex socket head screws [21] on the top of each
diaphragm housing with 5/32 hex key.
Figure G
17. Using 5/32 in. long hex wrench, unscrew and
remove plug [22] from the bottom of the
manifold block.
18.
Remove rear plug [22] from the back of the
manifold block. Insert a long 5/32 hex key
through rear plug opening, unscrew and remove
the alarm bypass body [23], ball [24] and spring
[25].
Step 2: CLEANING
Precision Medical, Inc. recommends using an ultrasonic cleaner for cleaning all non-elastomeric and non-
metallic components. However, cleaning with an all-purpose liquid cleaner and rinsing with clean, warm water
may be substituted. Both methods require thoroughly blow drying all passages before reassembly. Follow the
ultrasonic cleaner manufacturer instruction.