beautypg.com

Example of set-up using a heliox cylinder, Warning – Precision Medical PM5400 Series User Manual

Page 9

background image

8

H

ELI

O

2

BLENDEr

HELIUM-OxygEn

WARNING

When pressurizing the Blender inlets, avoid pressure surges greater

than 100 psi (6.9 bar).

• Always use a heliox pressure reducing regulator set within 30-75 psi

(2.1-5.2 bar) when using heliox cylinders to supply gas to the Heliox

inlet of the Blender.

• Always use an oxygen pressure reducing regulator set within 30-75

psi (2.1-5.2 bar) when using oxygen cylinders to supply gas to the

Oxygen inlet of the Blender.

ExAmplE Of sET-Up UsINg A HElIOx CylINDER

NOTE:

Heliox regulator in use.

This manual is related to the following products: