Specifications – Precision Medical PM5400 Series User Manual
Page 5

4
H
ELI
O
2
BLENDEr
HELIUM-OxygEn
SPECIFICATIONS
High Flow
Low Flow
Model
PM5580: 80/20 heliox
PM5570: 70/30 heliox
PM5480: 80/20 heliox
PM5470: 70/30 heliox
Primary Outlet
Flow Range
15 - 120 l/min
3 - 30 l/min
With both supply pressures at 50 psi (3.4 bar)
with BLEED Closed
Auxiliary Outlet
Flow Range
2 - 100 l/min
0 - 30 l/min
With both supply pressures at 50 psi (3.4 bar)
with BLEED Open
Bleed Flow @
100%
F
I
O
2
13 l/min or less
at 50 psi (3.4 bar)
3 l/min or less
at 50 psi (3.4 bar)
Maximum
Combined
Flow (All Outlets)
≥
120 l/min
≥
30 l/min
Bypass Flow
(Loss of Helium or
Oxygen supply)
>
85 l/min
>
45 l/min
Bypass Alarm
Activation
50 psi
(3.45 bar)
60 psi
(4.14 bar)
50 psi
(3.45 bar)
60 psi
(4.14 bar)
13-25 psi
16-24 psi
18-22 psi
16-24 psi
0.9-1.7 bar 1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar
Alarm Reset:
When pressure differential is 6 psi (0.4 bar)
or less.
Alarm Sound Level:
≥
to 80 db at 1 ft (0.3 m)
Oxygen Concentration
Adjustment Range:
PM5480
PM5580
20 -
100%
PM5470
PM5570 30 - 100%
Gas Supply Pressure:
30 - 75 psi (2.1 - 5.2 bar) Heliox & Oxygen
within 10 psi (0.69 bar) of each other
Mixed Gas Stability:
±1% Oxygen
Connection Types:
DISS Type - Heliox & Oxygen Inlets & Outlets
and / or NIST Type - Heliox & Oxygen Inlets
Note: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).