Transport / storage requirements, Pressure drop, Dryness and composition for inlet gases – Precision Medical PM5400 Series User Manual
Page 6: Specifications, Dimensions, Depth, Width, Height, Weight, Shipping weight
5
H
ELI
O
2
BLENDEr
HELIUM-OxygEn
SPECIFICATIONS
continued
dimensions:
(without fittings)
Depth:
4.9 in
(12.5 cm)
Width:
2.3 in
(5.7 cm)
Height:
4.1 in
(10.4 cm)
Weight:
2.29 lbs
(1.04 kg)
Shipping Weight:
2.95 lbs
(1.34 kg)
Operating Temperature Range:
59°F to 104°F (15°C to 40°C)
Transport / Storage Requirements
Temperature Range:
-10°F to 140°F (-23°C to 60°C)
Humidity:
Max 95% Noncondensing
F
I
O
2
Accuracy:*
± 3% of full scale @ 50 psi (3.4 bar)
Pressure drop:
Low Flow:
≤
2 psi (0.14 bar) at inlet pressures from 30-90 psi
(2.1- 6.2 bar) and at 10 l/min flow rate at 60%
F
I
O
2
.
High Flow:
≤
3 psi (0.21 bar) at inlet pressures from 30-90 psi
(2.1- 6.2 bar) and at 30 l/min flow rate at 60%
F
I
O
2
.
The Helium-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.
The Helium-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 21647.
dryness and Composition for inlet gases:
Heliox:
Medical grade 80/20 or 70/30 is required.
Oxygen:
Oxygen supply must meet all requirements of USP
Medical Grade Oxygen.
dew Point:
(ONLY for CE
requirements)
Both inlets should remain 10°F (5.55°C) or more
below the lowest temperature to which the air
distribution system equipment is exposed. At a
temperature of 25°F (-3.9°C) and a pressure of
90 psi (6.33 kg/cm
2
) this equates to 2000 mg/m
3
.
* Accuracy of
F
I
O
2
will be affected if bleed flow is not engaged at low flows. (At or below 3 l/min for Low Flow and 15 l/min for High Flow).
* When Heliox tank pressure and oxygen outlet pressures are unbalanced, bleed may need to be engaged at a higher liter flow
to maintain accuracy.
Specifications are subject to change without prior notice.