Warning, Caution – Precision Medical PM5400 Series User Manual
Page 4

3
H
ELI
O
2
BLENDEr
HELIUM-OxygEn
WARNING
• Only trained, qualified medical personnel under the direct supervision of a
licensed physician should operate the
H
ELI
O
2
Blender.
• Use this
H
ELI
O
2
Blender only for its Intended Use as described in this
manual.
• Confirm prescribed dose before administering to patient. Monitor on a
frequent basis. Continuous Monitoring with an alarmed Oxygen Monitor
/ Analyzer is recommended.
• The
H
ELI
O
2
Blender shall be serviced by a qualified service technician.
• Always follow ANSI and CGA standards for Medical Gas Products,
Flowmeters and Oxygen Handling.
• The clinician must verify the Heliox tank concentration prior to utilizing
the Blender.
• An alarmed Oxygen Monitor / Analyzer must be used to verify oxygen
concentration.
• Accuracy of oxygen concentration will be affected if bleed is not activated
at flow settings below 15 l/min for the High Flow Blender, and 3 l/min for
the Low Flow Blender.
• dO NOT obstruct the alarm.
• dO NOT use Blender when alarm is sounding.
• dO NOT use oil in or around the Blender.
• dO NOT occlude or obstruct the bleed port on the auxiliary outlet of
the Blender.
• dO NOT use near any type of flame or flammable/explosive
substances, vapors or atmosphere.
• Oxygen Concentration dial does not rotate 360 degrees. Rotating the
dial beyond the endpoint settings will damage the Blender.
CAUTION
• Turn off gas supplies when
H
ELI
O
2
Blender is not in use.
• Store the
H
ELI
O
2
Blender in a clean, dry area when not in use.
• The
H
ELI
O
2
Blender contains magnetic, ferrous material that may
affect the results of an mRI.
• Ensure all connections are tight and leak free.
• Avoid excessive pressure surges greater than 100 psi (6.9 bar) when
pressuring the Blender inlets.
• DO NOT steam autoclave.
• DO NOT immerse
H
ELI
O
2
Blender into any liquid.
• DO NOT gas sterilize with (EtO) Ethylene Oxide.
• DO NOT use if dirt or contaminants are present on or around the
Blender or connecting devices.
• DO NOT smoke in an area where oxygen is being administered.
• DO NOT clean with aromatic hydrocarbons.
• Inlet pressure of device used in conjunction with the
H
ELI
O
2
Blender must
match inlet pressure of
F
I
O
2
Blender.
• When using a bottled high pressure gas source, always use a pressure
reducing regulator set within 30-75 psi (2.1-5.2 bar).