5

Breathing Frequency:

Up to 35 b/min

Cannula Requirement:

Maximum 7 foot (2.1 m) long standard

adult single lumen nasal cannula.

The effect on accuracy of flow due to variations in ambient temperature

is standard accuracy +7.3% @ 32°F (0°C) and -3.0% @ 104°F (+40°C) .

The 19MFA flowmeter models are calibrated at specified inlet pressures

as stated on Flowtube, 70°F (21°C), standard atmospheric pressure.
Operating Conditions:

Temperature:

35°F to 105°F (1.7°C to 40.6°C)

Altitude:

Sea level to 10,000 ft (0 to 3,048 m)

Storage Conditions:

Temperature: -40°F to 140°F (-40°C to 60°C)

Maximum Humidity: 95% Noncondensing

COMPONENT dESCRIPTION

CAUTION

Missing or illegible labels must be

replaced, contact Precision Medical, Inc.

Flow Control

Knob

Indicator

Pointer / Label

Thorpe

Tube

Oxygen

Inlet

Oxygen

Outlet

Selector Dial /

Settings Label

Product Serial #

/ Warning Label