PRINCIPLES OF OPERATION

The Precision Medical, Inc. EasyPulseFlow is a combination of two (2)

devices, a traditional Thorpe tube flowmeter and a pulse oxygen conserver.

The EasyPulseFlow is designed to be used with low pressure oxygen

systems at specified Inlet Pressure. It consists of a back pressure

compensated Thorpe tube flowmeter, a conserving module, and a selector

dial to choose between Continuous Flow mode or Pulse mode.

In the Continuous Flow mode the flowmeter operates similar to a traditional

Thorpe tube flowmeter: the flow is determined by setting the flow control

knob to the prescribed flow. In this mode it is capable of delivering metered

flows from 0 to 15 l/min and up to 24 l/min at flush flow.

In the Pulse mode the device operates as an oxygen conserver. Settings

of 1 through 5 are available and supply a similar F

IO

2

to the patient as

continuous flow. The conserving module controls the pulse size and

timing to the patient. It supplies a pulse of oxygen at the beginning of

each breath. This reduces the oxygen demand on the system and limits

the drying of the airways. The oxygen is supplied to the patient through

the nasal cannula.

SPECIFICATIONS

Inlet Pressure Range:

19MFA: 50 psi (3.4 bar)

19MFA AU: 4 bar (58.0 psi)

19MFA C

I

: 5 bar (72.5 psi)

Pulse Settings:

1, 2, 3, 4 and 5

(similar F

IO

2

to continuous flow value)

Continuous Flow Range:

0-15 l/min metered

(20-24 l/min max flush flow)

Pulse Accuracy:

Within ±15% of the nominal bolus

value (at each breath rate)

Continuous Flow Accuracy: ±0.25 l/min from 0.5 l/min up to 5 l/min

±0.5 l/min from 5 l/min up to 15 l/min

Savings Ratio:

Up to 5.7:1

Trigger Method:

Negative inspiratory effort from patient

inhalation