Troubleshooting – Precision Medical Conserving Flowmeter User Manual

Page 10

background image

9

TROUBLESHOOTING

If the EasyPulseFlow fails to function, consult the Troubleshooting
Guide below. If problem cannot be solved, consult your Provider.

Problem

Probable Cause

Remedy

No flow/
pulse

• Flowmeter not securely

attached to oxygen
outlet

• Thorpe tube flowmeter

not turned ON

• Selector Dial set

between settings

• Securely attach

flowmeter to oxygen
outlet

• Turn ON the Thorpe

tube flowmeter using
Flow Control Knob

• Position Selector Dial

at correct setting

Conserver
not sensing
breath

• Selector Dial set

between settings

• Cannula disconnected
• Nasal cannula kinked
• Humidity bottle in line

• Cannula greater than

7 ft long

• High flow cannula not

being used

• Position Selector Dial

at correct setting

• Connect cannula
• Replace cannula
• Remove humidity

bottle

• ONLY use a 7 ft or

less cannula

• Replace cannula with

a high flow cannula

Patient de-
saturated in
pulse mode

• Flowmeter set below

5 l/min in pulse mode

• Device not sensing

patient breath

• Increase flow ≥ 5 l/min

• Change to continuous

flow therapy