WARNING

• The EasyPulseFlow is not to be used in the Pulse mode by

patients who breathe through their mouths.

• dO NOT use oils, greases, organic lubricants or any combustible

materials on or near this product. Wash and dry hands properly

prior to usage.

• dO NOT use a humidifier when administering Oxygen therapy

on pulse settings.

• dO NOT use pulse settings while patient is sleeping unless

continuous pulse oximetry is utilized.

• dO NOT smoke in an area where oxygen is being administered.

• The EasyPulseFlow is designed to operate with a single

lumen, adult cannula with a maximum length of 7 foot (2.1 m).

• Flowmeter must meet specified inlet pressure range, see

“SPECIFICATIONS”.

• To use Pulse, Flow Control Knob must be set to 5 l/min or

greater.

CAUTION

• Only personnel instructed and trained in its use should operate

the EasyPulseFlow.

• The EasyPulseFlow contains magnetic, ferrous material that may

affect the results of an MRI.

• dO NOT autoclave.
• dO NOT gas sterilize with EtO (Ethylene Oxide).
• dO NOT clean with aromatic hydrocarbons.
• Avoid dropping the EasyPulseFlow or placing it in a position

where it could fall and become damaged.

• The EasyPulseFlow may not be able to detect all respiratory

efforts of the patient. (Shallow breathers may not be able to
trigger the EasyPulseFlow.)

• Operating the EasyPulseFlow outside its range of operating

conditions may affect its accuracy and performance.