WARNING

Read this User Manual before installing or operating the Flowmeter.

CAUTION

Inspect the Flowmeter for visual damage before use, DO NOT USE

if damaged.

OPERATING INSTRUCTIONS

NOTE: Precision Medical, Inc. strongly recommends the use of kink

proof Cannula.

1. Turn Knob to the “OFF” position.
2. Connect the Flowmeter to the appropriate gas source. The

appropriate gas and pressure are specified on the Flow Tube or

Flowmeter body.

3. Verify that the Float Ball is at the very bottom of the Flow Tube.

NOTE: If the Float is not resting at the bottom of the Flow Tube, the

product is leaking; consult the “TROUBLESHOOTING” Guide.

4. Adjust Flow:

To

increase - Turn Knob counterclockwise

To

decrease - Turn Knob clockwise

5. Set flow by aligning center of Float Ball with indicator lines on the Flow

Tube.

6. Adjusting flow beyond the last calibrated indicator line will result in

an undetermined flow.

7. To obtain maximum flush flow, turn Knob fully Counterclockwise.

NOTE: Flush flow is any flow above the last calibrated line on the

Flow Tube with an unrestricted flow, as indicated on flowmeter

labeling.