Specifications – Precision Medical Chrome Flowmeters User Manual
Page 4

SPECIFICATIONS
Flow Range
Graduations
Accuracy
0 – 200 cc
20 cc (0-200)
0-100 cc ±10 cc
101-200 cc ±14 cc
0 – 1 l/min
.1 (0-1) l/min
0-1 ±.05 l/min
0 – 3.5 l/min
.125 (0-1) l/min
.25 (1-3.5) l/min
0-3.5 ±.15 l/min
0 – 5 l/min
.25 (0-5) l/min
0-5 ±.20 l/min
0 – 6 l/min
.5 (0-6) l/min
0-6 ±.50 l/min
0 – 8 l/min
.5 (0-8) l/min
0-8 ±.25 l/min
0 – 15 l/min
.5 (0-5) l/min
1 (5-15) l/min
0-5 ±.25 l/min
6-15 ±.50 l/min
0 – 26 l/min
1 (2-26) l/min
2-4 ±.50 l/min
5-26 ±10% of reading
0 – 70 l/min
5 (0-70) l/min
0-70 ±10% of reading
For the most current manual revision
please visit our website: www.precisionmedical.com
Tell us how we are doing!
Visit us at www.precisionmedical.com
Flush Flow is the output of the flowmeter when the flow indicator is
beyond the highest calibrated graduation. The Flush Flow range is as
indicated on the flowmeter labeling.
Transport / Storage Requirements -40˚F (-40˚C) to 140˚F (60˚C)
The gas and inlet pressures are indicated on the Flow Tube or Flowmeter body.
NOTE: Storage / Transport outside the specified range may cause
damage to the flowmeter.
The effect on accuracy of flow due to variations in ambient temperature is
standard accuracy +7.3% @ 32°F (0°C) and -3.0% @ 104°F (40°C).
The above Flowmeter models are calibrated at specified inlet pressure, 70°F
(21°C), standard atmospheric pressure. International models are calibrated
per specifications marked on Flow Tube or Flowmeter body.
Specifications are subject to change without prior notice.