Specifications – Precision Medical Chrome Flowmeters User Manual

Page 4

background image

SPECIFICATIONS

Flow Range

Graduations

Accuracy

0 – 200 cc

20 cc (0-200)

0-100 cc ±10 cc

101-200 cc ±14 cc

0 – 1 l/min

.1 (0-1) l/min

0-1 ±.05 l/min

0 – 3.5 l/min

.125 (0-1) l/min

.25 (1-3.5) l/min

0-3.5 ±.15 l/min

0 – 5 l/min

.25 (0-5) l/min

0-5 ±.20 l/min

0 – 6 l/min

.5 (0-6) l/min

0-6 ±.50 l/min

0 – 8 l/min

.5 (0-8) l/min

0-8 ±.25 l/min

0 – 15 l/min

.5 (0-5) l/min

1 (5-15) l/min

0-5 ±.25 l/min

6-15 ±.50 l/min

0 – 26 l/min

1 (2-26) l/min

2-4 ±.50 l/min

5-26 ±10% of reading

0 – 70 l/min

5 (0-70) l/min

0-70 ±10% of reading

For the most current manual revision

please visit our website: www.precisionmedical.com

Tell us how we are doing!

Visit us at www.precisionmedical.com

Flush Flow is the output of the flowmeter when the flow indicator is

beyond the highest calibrated graduation. The Flush Flow range is as

indicated on the flowmeter labeling.
Transport / Storage Requirements -40˚F (-40˚C) to 140˚F (60˚C)
The gas and inlet pressures are indicated on the Flow Tube or Flowmeter body.

NOTE: Storage / Transport outside the specified range may cause

damage to the flowmeter.

The effect on accuracy of flow due to variations in ambient temperature is

standard accuracy +7.3% @ 32°F (0°C) and -3.0% @ 104°F (40°C).
The above Flowmeter models are calibrated at specified inlet pressure, 70°F

(21°C), standard atmospheric pressure. International models are calibrated

per specifications marked on Flow Tube or Flowmeter body.

Specifications are subject to change without prior notice.