Caution, Warning – Precision Medical Chrome Flowmeters User Manual
Page 3

CAUTION
• Flowmeters must be operated with the Flow Tube in a vertical,
upright position.
• Only personnel instructed and trained in its use should operate
this Flowmeter.
• Ensure all connections are tight and leak free.
• Only use oxygen-safe leak detector.
•
DO NOT autoclave.
•
DO NOT gas sterilize with EtO (Ethylene Oxide).
•
DO NOT clean with aromatic hydrocarbons.
•
DO NOT immerse Flowmeter in any kind of liquid. This will void
the warranty.
• The 1MFA3001, 4MFA1001 & 6MFA1001 Flowmeters may have a
factory installed restrictor. Prior to use, check Flowmeter labeling
for flow restrictions.
• The 1MFA3001, 4MFA1001 & 6MFA1001 Flowmeters contain a
glass Flow Tube which is fragile. Special care should be observed
to avoid breaking the Flow Tube.
(Continued on Inside)
WARNING
•
Use Flowmeters only for their “Intended Use” as described in this
manual.
• ALWAYS confirm prescribed flow before administering to patient
and monitor flow on a frequent basis.
• Flowmeters may contain magnetic, ferrous material that may
affect the results of an MRI.
To Reduce the Risk of Fire or Explosion:
•
ALWAYS follow ANSI and CGA standards for Medical Gas
Products and Flowmeters and Oxygen Handling.
•
DO NOT use or store oils, greases, organic lubricants or any
combustible materials on or near this Flowmeter.
•
DO NOT use near any type of flame or flammable/explosive
substances, vapors or atmosphere.
•
DO NOT smoke in an area where oxygen is being administered.