To replace Inline Hydrophobic Filter:

1. Disconnect filter assembly from Suction Canister and remove Suction

Canister from the Aspirator.

2. Unscrew filter assembly from the Aspirator.

3. Dispose of filter.

4. Install new filter (Part #506047) by screwing filter into the Aspirator, until hand

tight. Then reconnect tubing and elbow.

WARNING

• The Easy Go Vac Aspirator contains a sealed lead acid battery.

DO NOT

dispose with normal trash. Consult your local waste management authority

for proper disposal procedures.

• If Aspirator is used for treatment of infectious disease, consult your Physician

or Healthcare Professional for recommended procedures for proper disposal.

CAUTION

DO NOT use sealing material on the filter threads.

Tubing is single patient use only. DO NOT clean.

Please Recycle

DISPOSAL INSTRUCTIONS

Dispose of Bio-waste products as described by your Healthcare Professional or

Healthcare Facility protocol.

Cleaning for Exterior Components:

1. Disconnect the Aspirator from power source and other connections before

cleaning.

2. Use a cloth or sponge dampened with soap and water.

3. Wipe dry with a clean cloth or paper towel.

4. Store the Aspirator in a clean area free from grease, oil, and other sources of

contamination.

RETURNS

Returned products require a Returned Goods Authorization (RGA) number, contact

Precision Medical, Inc. All returns must be packaged in sealed containers to prevent

damage. Precision Medical, Inc. will not be responsible for goods damaged in

transit. Refer to Precision Medical, Inc. Return Policy available on the Internet,

www.precisionmedical.com.

Manuals available on our Website; www.precisionmedical.com