OPERATING INSTRUCTIONS

CAUTION

DO NOT run the Easy GO Vac Aspirator without the Inline Hydrophobic

Filter. This will void the warranty.

CAUTION

Inspect the Easy Go Vac Aspirator for visual damage, DO NOT USE if

damaged.

1. Place the Easy Go Vac Aspirator in the upright position on a level surface.
2. Make sure the On/Off ( / ) Switch is in the “OFF” position.

3. Be sure the Inline Hydrophobic Filter and tubing is properly connected to

the vacuum port on the Suction Canister lid.

• The Inline Hydrophobic Filter prevents liquid from entering the vacuum pump.

4. Select appropriate Mode of Operation.

A. AC ELECTRICAL RECEPTACLE OPERATION

(100 - 240 Volts AC)

a. Insert the small plug on the cord from the Battery Charger/Power Supply

into

External Power Jack on the Easy Go Vac Aspirator.

b. Connect the female end of the AC Power Cord into the Battery Charger/

Power Supply.

c. Plug the male end (3 prong plug) into a grounded receptacle.

5. Push the On/Off ( / ) Switch to the “ON” position.

6. Block the patient connection on the Suction Canister and adjust the

Vacuum Control Knob to the desired vacuum.

To INCREASE vacuum - Turn knob CLOCKWISE

To DECREASE vacuum - Turn knob COUNTERCLOCKWISE

NOTE: Vacuum should be set as prescribed by your Provider.

7. Connect one end of the Patient Suction Tubing to the port on the canister

marked Patient Tubing.

8. Connect the other end of the Patient Suction Tubing to the suction catheter

and begin procedure as instructed by your Provider.

9. When procedure is completed, press the On/Off ( / ) Switch to the

“OFF” position.

NOTE: DO NOT overfill the Canister.