TROUBLESHOOTING
If the Easy Go Vac Aspirator fails to function, consult the Troubleshooting Guide.
If the problem cannot be corrected, consult your Provider.
There are no user serviceable parts inside. DO NOT open case.
Problem
Probable Cause
Remedy
Aspirator will not
operate on battery
1. On/Off Switch in “OFF”
position
2. Battery not charged
(Low battery light on)
1. On/Off Switch to “ON’’
position
2. •Recharge battery
•If battery will not charge,
contact your Provider
Aspirator will not
operate when
connected to AC
Power Source
1. On/Off Switch in “OFF”
position
2. AC Power Cord not
connected
1. On/Off Switch to “ON”
position
2. Check connection at
both AC Power Source
and Aspirator
Aspirator will not
operate when
connected to Auto
DC Power Cord
1. On/Off Switch in “OFF”
position
2. Auto DC Power Cord not
connected properly
3. Defective auto power port
receptacle
4. Defective Auto DC Power
Cord
1. On/Off Switch to “ON”
position
2. Check connections at
both ends of cord
3. Connect to a known
working auto power port
4. Contact your Provider
Battery will not
charge
1. Not connected to Battery
Charger / Power Supply
2. Defective Battery Charger
/ Power Supply
3. Defective battery
1. Connect Battery
Charger/Power Supply
2. Contact your Provider
3. Contact your Provider
Unable to adjust
vacuum
1. Filter assembly not
connected properly
2. Suction Canister not
connected properly
3. Filter clogged
4. Defective Suction Canister
5. Defective Aspirator
1. Confirm and tighten
connections
2. Confirm and tighten
connections
3. Replace filter assembly
4. Replace Suction Canister
5. Contact your Provider