TROUBLESHOOTING

If the Easy Go Vac Aspirator fails to function, consult the Troubleshooting Guide.

If the problem cannot be corrected, consult your Provider.

There are no user serviceable parts inside. DO NOT open case.

Problem

Probable Cause

Remedy

Aspirator will not

operate on battery

1. On/Off Switch in “OFF”

position

2. Battery not charged

(Low battery light on)

1. On/Off Switch to “ON’’

position

2. •Recharge battery

•If battery will not charge,

contact your Provider

Aspirator will not

operate when

connected to AC

Power Source

1. On/Off Switch in “OFF”

position

2. AC Power Cord not

connected

1. On/Off Switch to “ON”

position

2. Check connection at

both AC Power Source

and Aspirator

Aspirator will not

operate when

connected to Auto

DC Power Cord

1. On/Off Switch in “OFF”

position

2. Auto DC Power Cord not

connected properly

3. Defective auto power port

receptacle

4. Defective Auto DC Power

Cord

1. On/Off Switch to “ON”

position

2. Check connections at

both ends of cord

3. Connect to a known

working auto power port

4. Contact your Provider

Battery will not

charge

1. Not connected to Battery

Charger / Power Supply

2. Defective Battery Charger

/ Power Supply

3. Defective battery

1. Connect Battery

Charger/Power Supply

2. Contact your Provider

3. Contact your Provider

Unable to adjust

vacuum

1. Filter assembly not

connected properly

2. Suction Canister not

connected properly

3. Filter clogged

4. Defective Suction Canister

5. Defective Aspirator

1. Confirm and tighten

connections

2. Confirm and tighten

connections

3. Replace filter assembly

4. Replace Suction Canister

5. Contact your Provider