Warning, Cleaning / maintenance charging battery, Caution – Precision Medical PM65 EasyGoVac Aspirator User Manual
Page 8: Inline hydrophobic filter
7
5. Push the On/Off ( / ) Switch to the “ON” position. If yellow Low Battery
Light stays on continuously, battery needs charging. Charge battery or use
another mode of operation.
WARNING
• Turn the Aspirator “OFF” and disconnect from any external power
source before cleaning.
• DO NOT clean the Aspirator with Alcohol, Acetone or other solvents.
CLEANING / MAINTENANCE
CHARGING BATTERY
1. Insert small plug on cord from the Battery Charger/Power Supply into
External Power Jack on the Aspirator.
2. Connect the female end of AC Power Cord into the Battery Charger/Power Supply.
3. Plug male end (3 prong plug) into a grounded receptacle.
Amber light on
Battery Charger / Power Supply will come on. When amber light changes to
green, battery is fully charged. Average charging time is 2-4 hours.
(A new battery may take longer to fully charge the first time.)
NOTE: No damage will occur if the charger is left connected.
4. Disconnect AC Power Cord from AC receptacle.
5. Disconnect cord from
External Power Jack on Aspirator. When fully
charged, battery will provide approximately 120 minutes of operation in 8,
15 minute intermittent duration periods.
WARNING
Operating Easy Go Vac Aspirator after yellow
Low Battery Light has
been on continuously, will cause automatic shut down.
CAUTION
DO NOT leave the Aspirator connected to the auto power port when not
in use. Failure to follow these instructions can cause severe battery failure
and will
void the warranty, see “CHARGING BATTERY” .
6. Steps 6 thru 11, see “OPERATING INSTRUCTIONS”.
Inline Hydrophobic Filter:
This filter cannot be cleaned. If it becomes contaminated or clogged, it must be replaced.