CLEANING / MAINTENANCE

WARNING

Turn the Easy Vac Aspirator off and disconnect from any external
power source before cleaning.

Inline Hydrophobic Filter:

This filter cannot be cleaned. If it becomes contaminated or clogged, it
must be replaced.
To replace Inline Hydrophobic Filter:
1. Disconnect filter assembly from Suction Canister and remove Suction

Canister from the Aspirator.

2. Remove filter assembly from the Aspirator.
3. Dispose of filter.
4. Install new filter (Part #1690) and reconnect tubing and elbow.
Tubing is single patient use only. DO NOT clean.

Cleaning for Exterior Components:

1. Disconnect the Aspirator from power source and other connections

before cleaning.

2. Use a cloth or sponge dampened with soap and water.
3. Wipe dry with a clean cloth or paper towel.
4. Store the Aspirator in a clean area free from grease, oil, and other

sources of contamination.