TROUBLESHOOTING

If the Easy Vac Aspirator fails to function, consult the Troubleshooting
Guide below. If the problem cannot be corrected, consult your Provider.

Problem

Probable Cause

Remedy

Aspirator will not

operate

1. On/Off switch in

“OFF” position

2. Power Cord not

connected properly

3. Thermal limit

exceeded

1. Push On/Off switch

to “ON” position

2. Check plug at outlet

3. Shut Aspirator off,

wait 15 minutes; be

sure vents are clear

NO vacuum

on gauge / NO

vacuum at Patient

Connection Tubing

1. Hydrophobic Filter

not connected

properly

2. Suction Canister

not connected or

assembled properly

3. Defective Suction

Canister

4. Defective Regulator

1. Confirm and tighten

connections

2. Confirm and tighten

3. Replace Suction

Canister

4. Contact your

Provider

NO vacuum on

gauge / vacuum

at Patient

Connection Tubing

1. Defective vacuum

gauge

1. Contact your

Provider

Vacuum on gauge

/ NO vacuum at

Patient Suction

Tubing

1. Clogged

Hydrophobic filter

2. Canister overfilled

1. Replace

Hydrophobic filter

2. Empty canister

Unable to adjust

vacuum

1. Defective regulator

1. Contact your

Provider