GROUNdING INSTRUCTIONS

The Easy Vac Aspirator should be grounded. In the event of an electrical

short circuit, grounding reduces the risk of electrical shock by providing

an escape wire for the electrical current. This Aspirator is equipped

with a cord having a grounding wire with a grounding plug. The plug

must be plugged into an outlet that is properly installed and grounded.

dANGER

Improper use of the grounding plug can result in a risk of

electrical shock.

WARNING

Ground reliability can only be achieved when the power cord is
connected to an equivalent receptacle marked "Hospital Grade".

OPERATING INSTRUCTIONS

CAUTION

Inspect the Easy Vac Aspirator for visual damage, DO NOT USE
if damaged.

1. Place Easy Vac Aspirator near a properly grounded electrical

receptacle on a table, floor or other horizontal, smooth, hard surface.

Allow a minimum of 4 in. (10 cm) clearance on all sides to ensure

adequate air flow for cooling of Aspirator.

2. Press down on Aspirator until Suction Feet are secure.
3. Be sure the On/Off Switch is in the “OFF” position.
4. Be sure the Inline Hydrophobic Filter is properly connected to the

vacuum port on the Suction Canister lid.

• The Inline Hydrophobic Filter prevents liquid from entering

the vacuum pump.