Grounding instructions, Danger, Warning – Precision Medical PM60 EasyVac Aspirator User Manual
Page 6: Operating instructions caution
5
GROUNdING INSTRUCTIONS
The Easy Vac Aspirator should be grounded. In the event of an electrical
short circuit, grounding reduces the risk of electrical shock by providing
an escape wire for the electrical current. This Aspirator is equipped
with a cord having a grounding wire with a grounding plug. The plug
must be plugged into an outlet that is properly installed and grounded.
dANGER
Improper use of the grounding plug can result in a risk of
electrical shock.
WARNING
Ground reliability can only be achieved when the power cord is
connected to an equivalent receptacle marked "Hospital Grade".
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Vac Aspirator for visual damage, DO NOT USE
if damaged.
1. Place Easy Vac Aspirator near a properly grounded electrical
receptacle on a table, floor or other horizontal, smooth, hard surface.
Allow a minimum of 4 in. (10 cm) clearance on all sides to ensure
adequate air flow for cooling of Aspirator.
2. Press down on Aspirator until Suction Feet are secure.
3. Be sure the On/Off Switch is in the “OFF” position.
4. Be sure the Inline Hydrophobic Filter is properly connected to the
vacuum port on the Suction Canister lid.
• The Inline Hydrophobic Filter prevents liquid from entering
the vacuum pump.