Caution – Precision Medical PM60 EasyVac Aspirator User Manual
Page 7
6
CAUTION
dO NOT run the Easy Vac Aspirator without Inline
Hydrophobic Filter. This will void the warranty.
5. Plug the Aspirator into a properly grounded receptacle.
6. Turn the On/Off Switch to the “ON” position.
7. Block the patient connection on the Suction Canister and adjust
the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• Vacuum should be set as prescribed by Provider.
8. Connect one end of the Patient Suction Tubing to the port on the
Canister marked Patient Tubing.
9. Connect the other end of the Patient Suction Tubing to the
suction catheter and begin procedure as instructed by Provider.
10. When procedure is completed, press the On/Off Switch to the
“OFF” position.
• DO NOT overfill the canister
11. Empty the Suction Canister as follows:
• Remove the canister from its bracket and empty as instructed
by your Provider.
CAUTION
Canister is for single patient use only.
12. Replace Suction Canister and reconnect all suctioning tubing.