Caution – Precision Medical PM60 EasyVac Aspirator User Manual

Page 7

background image

6

CAUTION

dO NOT run the Easy Vac Aspirator without Inline
Hydrophobic Filter. This will void the warranty.

5. Plug the Aspirator into a properly grounded receptacle.
6. Turn the On/Off Switch to the “ON” position.
7. Block the patient connection on the Suction Canister and adjust

the Vacuum Control Knob to the desired vacuum.

To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE

• Vacuum should be set as prescribed by Provider.

8. Connect one end of the Patient Suction Tubing to the port on the

Canister marked Patient Tubing.

9. Connect the other end of the Patient Suction Tubing to the

suction catheter and begin procedure as instructed by Provider.

10. When procedure is completed, press the On/Off Switch to the

“OFF position.
• DO NOT overfill the canister

11. Empty the Suction Canister as follows:

• Remove the canister from its bracket and empty as instructed

by your Provider.

CAUTION

Canister is for single patient use only.

12. Replace Suction Canister and reconnect all suctioning tubing.