Transport / storage requirements, Pressure drop, Dryness and composition for inlet gases – Precision Medical Air-Oxygen Blender User Manual
Page 7

6
Blender
Air-Oxygen
SPECiFiCATiONS
continued
dimensions:
(without fittings)
Depth:
4.9 in
(12.5 cm)
Width:
2.3 in
(5.7 cm)
Height:
4.1 in
(10.4 cm)
Weight:
2.29 lbs
(1.04 kg)
Shipping Weight:
2.95 lbs
(1.34 kg)
Operating Temperature Range:
59°F to 104°F (15°C to 40°C)
Transport / Storage Requirements
Temperature Range:
-10°F to 140°F (-23°C to 60°C)
Humidity:
max 95% Noncondensing
F
I
O
2
Accuracy:*
± 3% of full scale
Pressure drop:
Low Flow:
≤ 2 psi (0.14 bar) at inlet pressures from 30-90 psi
(2.1- 6.2 bar) and at 10
l/min
flow rate at 60%
F
I
O
2
.
High Flow:
≤ 3 psi (0.21 bar) at inlet pressures from 30-90 psi
(2.1- 6.2 bar) and at 30
l/min
flow rate at 60%
F
I
O
2
.
The Air-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.
The Air-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.
The Oxygen Analyzer should comply with ISO 21647.
dryness and Composition for inlet gases:
Air:
medical Air supply should meet the requirements of
ANSI Z86.1 - 1973 commodity specification for Air, type
1 grade D or better.
Oxygen:
Oxygen supply must meet all requirements of USP
medical Grade Oxygen.
dew Point:
(ONLY for CE
requirements)
Both inlets should remain 10°F (5.55°C) or more
below the lowest temperature to which the air
distribution system equipment is exposed. At a
temperature of 25°F (-3.9°C) and a pressure of
90 psi (6.33 kg/cm
2
) this equates to 2000 mg/m
3
.
* Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 l/min for the high flow
Blender, and 3 l/min for the low flow Blender.
Specifications are subject to change without prior notice.