Blender

Air-Oxygen

OPERATiNg iNSTRUCTiONS

CAUTiON

Inspect the Air-Oxygen Blender for visual damage before

use, DO NOT USE if damaged.

1. Secure Blender to wall or pole mount bracket.

2. Connect Air and Oxygen supply lines from Blender to wall

outlets.

3. Connect flowmeter to Blender outlet.

4. Adjust the Oxygen Concentration Dial to the prescribed

concentration.

NOTE: The Oxygen Concentration Dial does not rotate 360°.

dO NOT force dial less than 21% or over 100% oxygen,

this will damage the Blender.

5. Confirm the flow of air and/or oxygen mixture to the patient.
6. Confirm the concentration of air/oxygen with an Oxygen

Analyzer/Monitor. If necessary activate bleed flow valve to

maintain F

I

O

2

accuracy.

7. To activate the bleed, turn and rotate the knurled collar back

until it contacts the cover.

8. To deactivate the bleed, pull and rotate the collar away from

the cover until bleed flow valve is closed.

9. Turn “OFF” gas supplies when Air-Oxygen Blender is not in use.

CLEANiNg

CAUTiON

dO NOT steam autoclave.

dO NOT immerse the Air-Oxygen Blender into any liquid.

dO NOT use any strong solvent or abrasive cleaners.

dO NOT gas sterilize with (EtO) Ethylene Oxide.

dO NOT clean with aromatic hydrocarbons.

1. Disconnect all gas connections and equipment before

cleaning.

2. Clean exterior surfaces with a cloth dampened with mild

detergent and water.

3. Wipe dry with a clean cloth.