Operating instructions, Cleaning – Precision Medical Air-Oxygen Blender User Manual
Page 12

11
Blender
Air-Oxygen
OPERATiNg iNSTRUCTiONS
CAUTiON
Inspect the Air-Oxygen Blender for visual damage before
use, DO NOT USE if damaged.
1. Secure Blender to wall or pole mount bracket.
2. Connect Air and Oxygen supply lines from Blender to wall
outlets.
3. Connect flowmeter to Blender outlet.
4. Adjust the Oxygen Concentration Dial to the prescribed
concentration.
NOTE: The Oxygen Concentration Dial does not rotate 360°.
dO NOT force dial less than 21% or over 100% oxygen,
this will damage the Blender.
5. Confirm the flow of air and/or oxygen mixture to the patient.
6. Confirm the concentration of air/oxygen with an Oxygen
Analyzer/Monitor. If necessary activate bleed flow valve to
maintain F
I
O
2
accuracy.
7. To activate the bleed, turn and rotate the knurled collar back
until it contacts the cover.
8. To deactivate the bleed, pull and rotate the collar away from
the cover until bleed flow valve is closed.
9. Turn “OFF” gas supplies when Air-Oxygen Blender is not in use.
CLEANiNg
CAUTiON
•
dO NOT steam autoclave.
•
dO NOT immerse the Air-Oxygen Blender into any liquid.
•
dO NOT use any strong solvent or abrasive cleaners.
•
dO NOT gas sterilize with (EtO) Ethylene Oxide.
•
dO NOT clean with aromatic hydrocarbons.
1. Disconnect all gas connections and equipment before
cleaning.
2. Clean exterior surfaces with a cloth dampened with mild
detergent and water.
3. Wipe dry with a clean cloth.