TROUBLESHOOTiNg

If the Air-Oxygen Blender fails to function, consult the

Troubleshooting Guide below. If problem cannot be solved by

using Troubleshooting Guide, refer to the Air-Oxygen Blender

Service manual (P/N 504827) available on the Internet;

www.precisionmedical.com or consult your Provider.

Problem

Probable Cause

Remedy

Oxygen

concentration

discrepancy

between Blender

setting and

Analyzer/Monitor

(greater than 3%)

1. High Flow model,

flow requirement

below 15 l/min.

Low Flow model,

flow requirement

below 3 l/min.

2. Analyzer/monitor

inaccurate

3. Low flow bleed

obstructed

4. Gas supply

contaminated

5. Downstream device

causing back flow or

restricted flow

1. Use auxiliary outlet

& engage bleed

2. Recalibrate

Analyzer/monitor or

Verify with second

Analyzer/monitor

3. Remove obstruction

4. Check gas sources

with calibrated

Oxygen Analyzer/

monitor to

confirm

Oxygen is 100%

and Air is 21%

5. Isolate Blender.

Check oxygen

concentration at

Blender Outlets

No flow at

Blender outlets

1. Gas sources turned

“OFF”

2. Gas sources not

connected

1. Turn gas sources

“ON”

2. Connect gas

sources

Alarm sounding 1. Difference between

Oxygen and air inlet

pressures greater

than specified

1. Correct pressure

difference until

Air and Oxygen

pressures are within

specification