Troubleshooting – Precision Medical Air-Oxygen Blender User Manual
Page 14

13
Blender
Air-Oxygen
TROUBLESHOOTiNg
If the Air-Oxygen Blender fails to function, consult the
Troubleshooting Guide below. If problem cannot be solved by
using Troubleshooting Guide, refer to the Air-Oxygen Blender
Service manual (P/N 504827) available on the Internet;
www.precisionmedical.com or consult your Provider.
Problem
Probable Cause
Remedy
Oxygen
concentration
discrepancy
between Blender
setting and
Analyzer/Monitor
(greater than 3%)
1. High Flow model,
flow requirement
below 15 l/min.
Low Flow model,
flow requirement
below 3 l/min.
2. Analyzer/monitor
inaccurate
3. Low flow bleed
obstructed
4. Gas supply
contaminated
5. Downstream device
causing back flow or
restricted flow
1. Use auxiliary outlet
& engage bleed
2. Recalibrate
Analyzer/monitor or
Verify with second
Analyzer/monitor
3. Remove obstruction
4. Check gas sources
with calibrated
Oxygen Analyzer/
monitor to
confirm
Oxygen is 100%
and Air is 21%
5. Isolate Blender.
Check oxygen
concentration at
Blender Outlets
No flow at
Blender outlets
1. Gas sources turned
“OFF”
2. Gas sources not
connected
1. Turn gas sources
“ON”
2. Connect gas
sources
Alarm sounding 1. Difference between
Oxygen and air inlet
pressures greater
than specified
1. Correct pressure
difference until
Air and Oxygen
pressures are within
specification