Specifications – Precision Medical Air-Oxygen Blender User Manual

Page 6

background image

5

Blender

Air-Oxygen

SPECiFiCATiONS

Model

PM5200 High Flow

PM5300 Low Flow

Primary Outlet

Flow Range

15 - 120

l/min

3 - 30

l/min

With both supply pressures at 50 psi (3.4 bar)

with BLEED closed

Auxiliary Outlet

Flow Range

2 - 100

l/min

0 - 30

l/min

With both supply pressures at 50 psi (3.4 bar)

with BLEED open

Bleed Flow

13

l/min

or less

at 50 psi (3.4 bar)

3

l/min

or less

at 50 psi (3.4 bar)

Maximum

Combined

Flow

(All Outlets)

120

l/min

30

l/min

Bypass Flow

(Loss of Air or

Oxygen supply)

>

85

l/min

>

45

l/min

Bypass Alarm

Activation

50 psi

(3.45 bar)

60 psi

(4.14 bar)

50 psi

(3.45 bar)

60 psi

(4.14 bar)

13-25 psi 16-24 psi 18-22 psi 16-24 psi

0.9-1.7 bar 1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar

Alarm Reset:

When pressure differential is 6 psi

(0.4 bar) or less.

Alarm Sound Level:

to 80 db at 1 ft (0.3 m)

Oxygen Concentration

Adjustment Range:

21 - 100%

gas Supply Pressure:

30 - 75 psi (2.1 - 5.2 bar)

Air and Oxygen within 10 psi (0.69 bar)

of each other

Mixed gas Stability:

±1% Oxygen

Connection Types:

DISS Type - Air & Oxygen Inlets & Outlets

and / or NIST Type - Air & Oxygen Inlets

Note: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).