Specifications, General – Precision Digital PD6770 User Manual
Page 5
![background image](https://www.manualsdir.com/files/858282/content/doc005.png)
PD6770 Loop-Powered Process Meter
Instruction Manual
5
SPECIFICATIONS
Except where noted all specifications apply to operation at +25°C.
General
DISPLAY
3 ½ digit LCD
1" (25.4 mm); -1999 to 1999
DISPLAY
UPDATE RATE
2.5 Updates/Second
OVERRANGE
Display reads
1
on the left most digit
PROGRAMMING
METHOD
Four front mounted rotary control dials accessed when the
cover is removed.
RECALIBRATION
Recalibration is recommended at least every 12 months.
NORMAL MODE
REJECTION
60 dB rejection ratio
ENVIRONMENTAL
Operating temperature range: -40 to 75°C
Storage temperature range: -40 to 75°C
Relative humidity: 0 to 90% non-condensing
CONNECTIONS
Screw terminals accept 12 to 22 AWG wire
ENCLOSURE
NEMA 4X, IP65 plastic field enclosure.
Color: grey.
Three ¾" NPT threaded conduit openings. One ¾" NPT
plastic conduit plug, with 1.29" wrenching flats and a
screwdriver slot, is included.
MOUNTING
May be mounted directly to conduit. Two slotted flanges for
wall mounting or NPS 1½" to 2½" or DN 40 to 65 mm pipe
mounting. See Mounting Dimensions on page 15.
OVERALL
DIMENSIONS
5.67" x 5.25" x 4.18" (W x H x D)
(144 mm x 133 mm x 106 mm)
WEIGHT
1.65 lbs (26.4 oz, 0.75 kg)
WARRANTY
3 years parts and labor