Epson Digital Imaging System EPIX User Manual
Page 19

Logos Imaging EPIX Scanner
User Manual
18
SECTION 7: MAINTENANCE PROCEDURES
The EPIX Digital Imaging System is designed for many years of trouble-free operation. It is
manufactured from the highest quality components to ensure excellent performance.
Maintenance that can be performed by the operator is minimal.
7.1. CLEANING THE SYSTEM
Cleaning the Scanner
Turn off the EPIX SCANNER, as described in
Section 5.5, before cleaning. Wipe the outside
surfaces with a paper towel dampened with a
cold sterilant solution or household, non-
abrasive cleaner (window cleaner works well).
DO NOT SPRAY OR SOAK THE
SCANNER. Wipe the inside of the carousel
well with a very slightly damp cloth. Be
careful not to allow running or dripping
solvents into the EPIX SCANNER. This could
cause damage to the electronics inside. Allow
to air dry before plugging in or turning back
on.
Cleaning Imaging Plates
Imaging plates should be handled carefully.
For the best images, take care not to scratch
them and keep them dust- free. Use the
following procedure to clean them:
1. Use lint- free, 100% cotton gauze (not
cotton balls). Gently wipe the cotton gauze
over the dry imaging plate surface. Wipe
back and forth and then in a circular
motion.
2. To clean any remaining stains, dampen the
gauze in anhydrous isopropyl alcohol and
wipe using the same motion as above.
3. Completely dry the surface by wiping with
another piece of cotton gauze. Ensure that
the imaging plate is completely dry before
use.
Cleaning the carousel:
Fig. 7-1
Spray the soiled surfaces of the imaging plate holder
and carousel with a general household non-abrasive
cleaner or soap and water. Do not use strong alkaline or
ammonia based cleaners. Wipe clean and allow to air
dry.