beautypg.com

Housings, Applications, Housings applications – Emerson Transmitters and Discrete Controllers Series 3000 User Manual

Page 4

background image

4

Micro Motion

®

Series 3000 Transmitters and Discrete Controllers

Housings

Applications

Rack-mount

19-inch (486.2 mm) rack-mount housing meets DIN standard 41494 and IEC 297-3

Front panel is rated IP40

Panel-mount

Front panel with bezel is rated NEMA 4X (IP65)

Field-mount

Compartment with threaded cover contains electronics

Terminal compartment contains:

Non-intrinsically safe compartment with input/output and power supply terminals

Intrinsically safe compartment with interface/display, sensor terminals (Model 3700 only), and
optional input and output terminals

Mounting bracket and interface/display rotate to allow mounting in four different orientations

Rated NEMA 4X (IP67)

Process monitor/totalizer

Process monitoring and totalizing for multiple variables, including mass flow, volume flow, density,
and temperature.

Discrete batch control

• Configuration of up to six preset batch recipes
• One-stage or two-stage batch control
• End-of-batch warning and batch-overrun alarms
• Configuration by quantity or by percent of target for primary valve closure, secondary valve

opening and closure, end warning, and batch overrun

• Configuration for target lockout, or for target adjustment during batch operation
• Automatic overshoot compensation

Meter verification

Provides

in-situ

assessment of a Micro Motion Coriolis meter, determining whether the meter has

been affected by erosion, corrosion, or other influences affecting meter calibration. No secondary
references are required to perform this operation.

Enhanced density

Enhanced density measurement for a number of applications, including:
• %HFCS, °Brix, °Plato, °Balling,
• °Baumé at SG60/60
• Density at reference temperature
• Specific gravity
• Concentration derived from reference density
• Concentration derived from specific gravity

Custody transfer

• Physical and software security
• Security-alarm posting
• Mass or volume totalizer that can be configured by the user
• Custom formatting of receipt tickets
• Audit trail of configuration changes
• Multidrop printing capability

Petroleum measurement (API) Provides API process variables, such as API volume flow and API average density.