beautypg.com

Model 3100 accessory, Model 3500/3700 liquid flow performance – Emerson Transmitters and Discrete Controllers Series 3000 User Manual

Page 10

background image

10

Micro Motion

®

Series 3000 Transmitters and Discrete Controllers

Model 3100 accessory

Model 3500/3700 liquid flow performance

Relays module

Three relays

Solid state, SPST

Rated at 24–250 VAC, 40 mA to 5 A; or 0–70 VDC, 5 A

Energized by discrete outputs

NEMA 4X (IP65) enclosure (optional)

UL and CSA approvals

Class I, Div. 2, Groups A, B, C, and D

ATEX approvals

Suitable for installation in Zone 2, when properly installed in a suitable enclosure, complying with
ATEX Directive (94/9/EC) for Group II, Category 3G, according to CENELEC standard EN 50021
(1998) and marked as:

EEx n V II T4
KEMA 97 ATEX4940X

Ambient temperature –20 to +60 °C (–4 to +140 °F)

Additional installation-specific requirements are described in detail in the Series 3000 instruction
manual.

Sensor model

Mass flow accuracy

(1)(2)

(1) Stated flow accuracy includes the combined effects of repeatability, linearity, and hysteresis.

(2) For the details of flow accuracy and repeatability specifications, refer to the product data sheet for each sensor family.

ELITE

®

±0.05%

of

rate

F-Series

±0.10% of rate

H-Series

±0.10% of rate

T-Series

±0.15% of rate

R-Series

±0.50% of rate

Volume flow accuracy

(1) (2)

ELITE

±0.05% of rate

F-Series

±0.15% of rate

H-Series

±0.15% of rate

T-Series

±0.25% of rate

R-Series

±0.50% of rate

Repeatability

(2)

ELITE

±0.025% of rate

F-Series

±0.05% of rate

H-Series

±0.05% of rate

T-Series

±0.05% of rate

R-Series

±0.25% of rate