beautypg.com

E&J GF0600059REVB06 User Manual

Page 35

background image

Metro Power

III Owner’s Manual

GF0600059RevB06, November 2006

35

SYMPTOM

BATTERY

POSSIBLE CAUSE

CORRECTIVE ACTION

GAUGE

LIGHTS

Wheelchair

Off

won’t drive

Flash fast

Control system fault

The controller safety circuits have operated and the controller has

(even with

been prevented from moving the wheelchair. This indicates that there

joystick

is a fault. Please follow this procedure:

released)

1. Turn the controller’s ON/OFF switch off.

2. Ensure that all controller and wheelchair connectors are securely mated.

3. Check battery condition.

4. If you can’t find the problem, try using the following Battery gauge diagnostic guide.

5. Turn the controller on again and try to drive the wheelchair. If the fault lights flash again,

do not try to use the wheelchair; contact your dealer.

Wheelchair

On, steady

All is well

None

drives

Flash

Low batteries

Controller is functioning correctly but you should charge the batteries

slowly

as soon as possible.

Wheelchair

On

Speed control or

If the wheelchair does not travel at full speed or does not respond

drives, but

nonhazardous fault

quickly enough, and battery condition is good, check speed control

movement is

setting. If adjusting the speed control does not remedy the problem,

sluggish

there may be a nonhazardous fault. Contact your dealer.

Speed and response

Reprogram to desired

have been

performance

programmed slow

Battery gauge diagnostic guide
In addition to the battery gauge function, the battery gauge provides a diagnostic display. When a

system fault occurs, a set (from one to ten) of light bars will flash. Use the illustration below to di-

agnose the fault (the number represents the number of light bars flashing), then find the problem in

the Battery gauge diagnostic guide table that follows.

Battery gauge diagnostic guide table

BATTERY

INDICATOR

SYSTEM

LIGHTS

FAULT

CHECK / ACTION

Red, yellow, &

None

Battery charged; controller and electrical system OK.

green bars lit


Red & yellow

None

Charge battery if possible; controller and electrical system OK.

bars lit


Red bars only

None

Low batteries; controller and electrical system OK. Charge batteries.

lit or slow flash


Red, yellow,

Possible

Joystick out of neutral—leave joystick in neutral while turning wheelchair power on. Operate

& green bars, joystick fault

joystick to make sure it returns to neutral. Turn ON/OFF switch off and ON to reset.

ripple up &

down