8

102R Pumphead

The 102R is a twin roller pumphead for high precision flow rates. It is limited to the use of silicone tubing only.

Flow rates

Flow rates f or the101F /R and 101U/R w ere obtained us ing W atson-Marlow 1.6m m w all thic kness Silic one tubing

pumping water at 20C with negligible suction and delivery pressures (unless otherwise stated). Where an application is

critical, the f low rate should be deter mined under operating conditions. The important factors are suction and delivery

pressures, temperature and fluid viscosity.

Tube loading

Isolate the drive from the mains supply.

Fit one end of the tubing into one of the s pring loaded c lamps , and then w hilst rotating the r otor by hand feed the

tubing ar ound the tr ack ( if r esistance is f elt to tur ning the r otor, tur n the r eversing s witch to the other direction of

rotation).

Fit the other end of the tubing into the s econd s pring loaded c lamp ens uring that the tubing is not slack in the

pumphead since this can reduce tube life.