Recommended operating procedures

DO

keep delivery and suction lines as short as possible using a minimum number of swept bends.

DO use suction and deliver y pipelines with a bor e equal to or larger than the bore of the tube fitted in the pumphead.

When pumping

viscous

fluids, the losses caused by increased friction can be overcome by using pipe runs with a cross

sectional area several times greater than the pumping element.
DO fit an ex tra length of pump tube in the s ystem to enable tube tr ansfer. This will extend tube lif e and m inimise the

downtime of the pumping circuit.
DO

keep the track and rollers clean.

The self-priming nature of peristaltic pumps means valves are not required. Any valves fitted must cause no restriction

to flow in the pumping circuit.
Tube selection T he chemical compatibility list published in the W atson-Marlow catalogue is only a guide. If in doubt

about the compatibility of a tube material with the duty fluid, request a tube sample card for immersion trials.

Installation

The 101F/R and 101U/R are suitable for single phase mains electricity supplies only.

To ensure correct lubrication of the gearbox the pum p should be run only while its feet are standing on a horizontal

surface. The pump should be positioned to allow a free flow of air around it.

Set the voltage selector to either 120V for 100-120V 50/60Hz supplies or 240V for 220-240V 50/60Hz supplies.
A mains cable fitted with a moulded plug is supplied with the pump. The wires are colour coded in accordance with the

following code:

220-240V: Live- Brown; Neutral - Blue; Earth - Green/Yellow.

100-120V: Live - Black; Neutral - White; Earth - Green.

Troubleshooting

Should the pump fail to operate, make the following checks to determine whether or not servicing is required.

Check that the power switch is on.

Check the mains supply is available at the pump.

Check the voltage selector switch is in the correct position.

Check the fuse in the mains socket.

Check that the pump is not stalled by incorrect fitting of tubing.

101F/R operation

Turn power switch to on (I) position to start pump.

Turn power switch to off (0) position to stop pump.

101U/R Manual operation

Set the front panel

Auto/Man

switch to

Man.

Turn power switch to on (I) position to start pump.

Turn power switch to off (0) position to stop pump.

Change pumping direction using the

CW/CCW switch on the front panel, and s et the pum p speed using the digital

potentiometer which is calibrated in percentage of maximum speed.

To prime the pump at maximum speed press the

Max key.

It is not nec essary to dis connect the pr ocess s ignal f rom the r ear panel 15-pin D connector or adjust the calibration

potentiometers if returning to manual control after the pump has been in automatic operation.