7 main menu, 8 troubleshooting, 9 pumphead replacement – Watson-Marlow Qdos30 Universal User Manual

Page 5: 2 wiring up the terminal connectors, 3 110 volt logic module pcb connectors, 1 general settings

background image

EN

Watson-Marlow qdos30 pumps product safety information

5

6.2

Wiring up the terminal connectors

Ensure that multi strand wires are terminated with a crimp, suitable for the wire

diameter. Failure to do so may result in electric shock

Ensure that unused openings on the module are sealed using the blanking plug

provided. Failure to do so may result in electric shock due to ingress of liquid or dust.

6.3

110 volt logic module pcb connectors

Apply the correct signals to the terminals shown below. Limit signals to the maximum

values shown. Do not apply voltage across other terminals. Permanent damage, not

covered by warranty, may result. The maximum load on the relay contacts of this

pump is 115VAC; max current 5A. Voltages do not require transformer isolation. The

relays are specified to switch 115VAC mains powered loads. These relay outputs are

not suitable for 24VDC logic signals.

7 Main menu

7.1

General settings

Do not use auto restart for more than 20 starts per hour. We recommend remote

control where a high number of starts is required.

8 Troubleshooting

If the pump display remains blank when the pump is switched on, make the following checks:
• Check that mains power is available to the pump.
• Check the fuse in the wall plug if one is present.
If the pump runs but there is little or no flow, make the following checks:
• Check that fluid is supplied to the pump.
• Check for any kinks or blockages in the lines.
• Check that any valves in the lines are open.

9 Pumphead replacement

Ensure contaminated pumpheads are not shipped but disposed of locally according to

regulations for contaminated items and health and safety procedures.

Always isolate the pump from the mains power supply before maintenance on the

pumphead, suction or discharge lines.

It is only possible to fit the pumphead in one orientation with the arrow pointing

upwards.

The pumphead retaining clamps should only be loosened or tightened by hand.