10. CONTROL OUTPUT POLARITY

Write Command:

2c hex

Read Command:

45 hex

Interpret:

0 is HEAT WP1+ and WP2-.

1 is HEAT WP2+ and WP1-.





11. POWER ON/OFF

Write Command:

2d hex

Read Command:

46 hex

Interpret:

0 is off

1 is on.


12. OUTPUT SHUTDOWN IF ALARM

Write Command:

2e hex

Read Command:

47 hex

Interpret:

0 is no shutdown upon alarming.

1 is to shutdown main output drive upon alarming.


13. FIXED DESIRED CONTROL SETTING

Write Command:

1c hex

Read Command:

50 hex

Interpret:

When writing, multiply the desired control temperature by 10 and
convert to hex. This becomes the “send value”
When reading, convert the “return value” to decimal and divide by
10.0 to convert to ° F/°C.


14. PROPORTIONAL BANDWIDTH

Write Command:

1d hex

Read Command:

51 hex

Interpret:

Fixed-point temperature bandwidth in °F/°C.

Multiply desired bandwidth by 10.

1 °F/°C bandwidth would be 10 decimal.

100 °F/°C bandwidth would be 1000.


15. INTEGRAL GAIN

Write Command:

1e hex

Read Command:

52 hex

Interpret:

Fixed-point gain in Repeats/min.

Multiply desired integral gain by 100.

0.01 rep/min would be 1 decimal.
1.00 rep/min would be 100 decimal.


16. DERIVATIVE GAIN

Write Command:

1f hex

Read Command:

53 hex