Warnings, Specifications – Super Systems 1500 VRSD User Manual
Page 5
Model 1500 VRSD Operations Manual
Super Systems Inc.
Page 5 of 28
Warnings
Although it is intended for use in an industrial environment, this is a sensitive piece of
analysis equipment. Care should be taken not to operate it in a manner inconsistent
with its intended use.
•
Moisture (water) cannot be allowed to enter the unit. If water is present in the
sample gas, use an in-line dryer for sample conditioning. In the event that the
sensor becomes wet, use an inert gas (Nitrogen or Argon) to dry the inside of the
instrument. Under no circumstances should Methane (Natural Gas) be used to dry
the unit.
•
The analyzer must be stored at ambient temperature (65-80°F) for at least four
hours prior to operation.
•
An in-line dryer for sample conditioning should be used for exothermic and
combustion applications.
•
This unit is not designed to measure the dew points in corrosive gasses, such as
Ammonia, S0
3
, Chlorine, and HCL.
IMPORTANT!
Failure to comply with these conditions may cause damage to the unit that will not be
covered under the warranty. Super Systems, Inc. is not responsible for damage to this
unit caused by disregard of these warnings, neglect, or misuse.
Specifications
Measurement Range:
0 to +80°F (-17.8 to +27°C)
Temperature Range:
0 to +120°F (-18 to +49°C)
Accuracy:
+/- 1°F (+/- 0.5°C)
Zero and Span Drift:
+/- 1°F (+/- 0.5°C)
Power Supply:
115 VAC 60Hz (Optional 220VDC)
Display Type:
3.5” Color Touch Screen
Display Resolution:
+/- 1°F (+/- 0.1°C)
Digital Communications:
(2) RS485, RS232, Ethernet
Control / Retransmission Output:
4 – 20 mA (2 Outputs)
Alarms:
Up to 3 Alarm Relays
Dimensions:
20” x 16” x 10” (508mm x 406mm x 254mm)
Weight:
34.4 lbs. (15.6 kg)