beautypg.com

Default settings, Modifying the default settings, Description of menu items – Super Systems Hydrogen Nitrider Analyzer User Manual

Page 5: Exit program, Instrument information, Communication setup, Instrument configuration, Percent hydrogen

background image

Basic Nitriding Sampling System – Hydrogen Analyzer

Super Systems Inc.

Page 5 of 12

Default Settings

When the instrument is turned on, it will display % H

2

on the screen. To display additional values (% DA,

% NH

3

, or K

N

), see the “Instrument Setup” screen. The first 4-20mA output will be set up to retransmit

the %H

2

value scaled for 0 to 100%, and the second output will be set up for %DA also scaled for 0-100%.

Any changes to these default parameters will be stored so they will not need to be re-entered when the

power is shut off to the instrument

Modifying the Default Settings

To access the operational and setup parameters, press the “menu” button at the lower left section of the

screen. This will allow you to select only two options, Exit Program and Instrument Information. To

prevent unwanted modification to the operation of the instrument, these are the only options available

unless the user logs in. To access the other menu options, press the “Login” button and enter “2”. This

will give the operator access to the setup and operational parameters.

Each individual menu option is described in detail below:

Description of Menu Items

Exit Program

The touch screen display is constantly writing data to its flash card for storage, and it is important that

the instrument not be shut down during this process. Before removing power from the enclosure, select

“Exit Program”. It will ask is you are sure that you want to shut down the interface. By answering yes, it

closes the operating program in an orderly manner. Once the standard Microsoft Windows screen

appears it is safe to remove power from the instrument. If this procedure is not followed, there is a

chance that there could be an error writing to the flash card that could cause problems with the operation

of the instrument.

Instrument Information

This screen provides information on any applicable revision levels and serial numbers. It also shows if

the instrument is logging data. There are no functions that can be performed on this screen; it is for

informational purposes only.

Communication Setup

The communication methods shown on this screen are for display only and cannot be modified. The baud

rates can be adjusted but they have been optimized for this instrument and modification is not

recommended.

Instrument Configuration

This instrument is capable of displaying four different parameters. These include:

Percent Hydrogen (H

2

)

Percent Dissociation (DA)

Percent Ammonia (NH

3

)

Nitriding Potential (K

N

)

The Percent Hydrogen is displayed on the main screen at all times. One additional parameter can be

displayed along with Hydrogen by highlighting it and pressing the “OK” button. A clearer description of

each of these parameters can be found in Appendix “A”, Gas Nitriding Technical paper.

Percent Hydrogen

The percent hydrogen is the amount of Hydrogen that is being detected by the thermal conductivity

sensor inside the instrument. There are no additional calculations being performed to this value.