Introduction, Specifications, Electrical connections / terminal assignments – Super Systems Hydrogen Nitrider Analyzer User Manual
Page 4: Getting started

Basic Nitriding Sampling System – Hydrogen Analyzer
Super Systems Inc.
Page 4 of 12
Introduction
This instrument uses the measurement of Hydrogen to display % Hydrogen (H
2
), % Dissociation (DA), and
% Ammonia (NH
3
). When the flow rates of Nitrogen, Ammonia, and Dissociated Ammonia are manually
entered, the instrument can also calculate Nitriding Potential (K
N
).
Specifications
•
Power Requirements: 100-240 VAC
•
Current Draw: Max. 0.2 Amps
•
Sensor Technology: Thermal Conductivity
•
User Interface: 3.5” Color QVGA TFT LCD Touch Screen
•
Measurement Range: 0-100% H
2
•
Hydrogen Measurement Accuracy: +/- 0.01%
•
Hydrogen Measurement Repeatability: +/- 0.01%
•
Hydrogen Measurement Resolution: 0.01%
•
Analog Outputs: Two Isolated 4-20mA (User Configurable)
•
Analog Output Resolution: 0.005mA
•
Analog Output Accuracy: +/- 0.01% of Range
•
Analog Output Linearity: +/- 0.01%
•
Analog Output Load Resistance: Minimum 0 Ohm, Maximum 500 Ohm
•
Digital Communications: RS485 Modbus, Ethernet
•
Enclosure Size (Without Filter): Approx. 13”L x 13”W x 6”D / 330mmL x 330mmW x 142mmD
•
Enclosure Weight: 22.5 lbs. / 10.2 kg.
•
Enclosure Ventilation: Continuous Purge Fan, Dual Vents
•
Recommended Flow Rate: 1.5 to 2.0 SCFH / 0.71 to 0.94 lpm
•
Process Gas Fittings: Stainless Steel Compression for ¼” OD Tubing
•
Calibration Gas Fitting: 1/8” Barb (can be removed for 1/8” Female NPT Port)
•
Operating Environment: 10-90 %RH (Non-Condensing)
•
Operating Temperature: 32 to 122°F / 0 to 50°C
•
Sample Gas Temperature: 32 to 158°F / 0 to 70°C
Electrical Connections / Terminal Assignments
Wire Number
Function
1000
AC Line Power (100-240VAC)
1002
AC Neutral Power
Ground
AC Ground
1121
Analog Output Common (+)
1111
Analog Output #1 (-)
1121
Analog Output Common (+)
1131
Analog Output #2 (-)
1081
RS485 (-)
1091
RS485 (+)
Getting Started
Please refer to the enclosed drawing for instructions regarding the proper electrical and mechanical
installation of this instrument. The flow meter on the door of the enclosure should be adjusted to
maintain 1.5 SCFH of process gas flow.