Super systems inc – Super Systems DPC2530 User Manual
Page 9

SSi
Super Systems Inc.
Product Manual #4553 - Model DPC2530 Continuous Dew Point Analyzer
SSi Manual 4553 – DPC2530
Super Systems, Inc.
Page # 9 of 17
Revision Level “E”
(800
) 666-4330
10.2
Turn the unit off.
11.0
Allow the sensor to reach equilibrium with the calibration salt.
11.1
With the power to the unit still turned off, leave the sensor in the
calibration salt for a minimum of twenty-four (24) hours. It is acceptable to
leave the sensor in the salt solution for a longer period of time, even a few
days, if desired.
12.0
Begin the 11.3% (Zero) calibration process
12.1
After leaving the sensor in the salt for at least twenty-four (24) hours, turn
the unit on. The reading on the display is not important at this point.
12.2
Simultaneously press the “11.3%” and “Calibration” buttons on the
microprocessor board.
13.0
Verify the 11.3% (Zero) calibration
13.1
Do not be concerned if the unit does not display 11.3, since it is not
supposed to match the value of the calibration salt.
13.2
Use the “Sensor Temp” switch on the face plate of the unit to determine
and record the sensor temperature.
13.3
Look up this temperature on the “Theoretical Dew Point Values for
Calibration Verification” chart located in the back of this manual. Appendix
“A” will show the temperature values in Fahrenheit, and Appendix B will
show the temperature values in Celsius.
13.4
Next to the appropriate temperature, note the number in the corresponding
column titled “11.3%”. This should match with the dew point that is shown
on the display of the DPC2530.
14.0
Determine the acceptability of the reading.
14.1
The value printed on the chart in Appendix A is a theoretical value, and
some variation can be expected. When a calibration is performed at SSi, we
certify (in writing) that the unit displays within +/- 1 degree of the theoretical
value after it has been calibrated. We would not consider a calibration to be