Product Manual #4553 - Model DPC2530 Continuous Dew Point Analyzer
SSi Manual 4553 – DPC2530
Super Systems, Inc.
Page # 3 of 17
Revision Level “E”
(800
) 666-4330
The analyzer must be stored at ambient temperature (65-80°F) for at least four hours
prior to operation.
An in-line dryer for sample conditioning should be used for exothermic and
combustion applications.
This unit is not designed to measure the dew points in corrosive gasses, such as
Ammonia, S0
3
, Chlorine, and HCL.
Please read and understand this Product Manual before operating the unit.
Failure to comply with these conditions may cause damage to the unit that will not be
covered under the warranty. Super Systems, Inc. is not responsible for damage to this
unit caused by disregard of these warnings, neglect, or misuse.
TERMINAL BLOCK WIRING –
Power and communications wiring for the DPC2530 should be performed according to
the following diagram:
INSTRUMENT SETUP –
The instrument should be fully configured at the factory for immediate use. In the event
that the settings are ever lost, the following chart shows the appropriate input parameters
for the 7EK controller. When the instrument is set up in Fahrenheit using these
parameters, the input is scaled from –50 to +80°F and the 4-20mA output is scaled from –
FU100 - 2.0 AMP
FU103 - 1.0 AMP
1000
Positive
1002
Negative
GND
Ground
1101
Shield
1111
Positive
1121
Negative
1151
Positive
1161
Negative
120V 1 PHASE 60 HZ
SUPPLY POWER
4-20 mA
OUTPUT
RS 485 (MODBUS)
COMMUNICATIONS