Introduction
The Model PGA3500 is a portable 3-Gas IR analyzer with an Oxygen (O
2
) cell on-board. It
measures Carbon Monoxide (CO), Carbon Dioxide (CO
2
) and Natural Gas (CH
4
) typically found
in an endothermic atmosphere.
Specifications
The unit is designed and manufactured for the atmosphere heat-treating industry.
CO range:
0.00 to 30.00 % *
CO
2
range:
0.000 to 2.000 %
CH
4
range:
0.00 to 15.00 % *
O
2
range:
0.1 to 25.0%
* Note: The CO and CH4 sensors have been calibrated to be most accurate within the ranges
as shown above, however both sensors are capable of measuring gas concentrations of up to
100.00%.
Sampling method:
Extraction by internal pump (when necessary)
Accuracy and repeatability:
± 1% of full scale
Flow Meter:
Inside case lid and also on-screen
Pump Operation:
On/Off/Automatic
AC Power:
90 to 230 VAC, 50 to 60 Hz, 60 Watts
DC Power:
12VDC rechargeable NiCd battery
(Charger integrated inside instrument)
Communications:
RS232, Ethernet, RS485 (Future Use)
Data
Storage:
Continuous
automatic
data
logging
Data Retrieval:
Visible in graphical and tabular format using
XGA Viewer software (included)
Operating Temperature:
32
° to 122° F (0° to 50° C)
Dimensions:
Approx. 16”H X 20”L X 8”D
Weight: Approx.
30
lbs.