3-Gas Analyzer PGA3000
Revision Level E
Menu List
The menu list shows the available pages displayed six at a time. The cursor up
(B) and the cursor down (C) keys are used to scroll the list.
OPERATOR MENU
1. IR STATUS DISPLAY
2. IR OVERALL STATUS DISPLAY
3. COMBUSTION DISPLAY
4. PUMP CONTROL
5. SET LCD DISPLAY (CONTRAST/BRIGHTNESS) VALUES
6. HELP
7. CALIBRATION DATES
11. ABOUT/SIGN-ON
12. START LOGGER
13. LOGGED DATA FILE
14. ZERO CALIBRATION
15. O2 CELL CALIBRATION
CONFIGURATION MENU – Pass code required
21. PORT SETUP
22. SET THE DATE AND TIME
23. DATA SET PAGE
24. SPAN CALIBRATION
25. SET PASS CODES
26. CLEAR LOGGED FILE
Pages number 8, 9, 10, 16, 17, 18, 19, 20 do not exist. The Configuration Menu
pages require the entry of a pass code to access them. The menu list page
shows the date and time at the bottom of the screen.
Pressing the # key allows the user to review the menus at anytime.
Screens 1 through 20 do not require any security codes. Screens 21 and higher
are configuration screens, and a pass code is required. The default pass code is
1, however this can be changed to any number between 0 and 512 by accessing
Menu Page 25.