20.2 In all measurements the P1 and P2 terminals are conventionally connected. Bear in

mind that thermal emfs may be present. The DO7e is particularly sensitive to these

on the 6m Ohm range.

20.3

For FS measurements, connect the C1 and C2 leads in the conventional manner (see

Fig

20.1).

20.4 For ZERO measurements, connect the C1 and C2 leads to the C2 terminal of the

standard (see Fig 20.2).

FULL SCALE CALIBRATION

P1, P2 LEADS - PURE COPPER, MULTI STRANDED

C1, C2 LEADS - 10A RATING (55/0.15MM)

Fig.

20.1

ZERO CALIBRATION

P1, P2 LEADS - PURE COPPER, MULTI STRANDED

C1, C2 LEADS - 10A RATING (55/0.15MM)

Fig.

20.2

* * END OF TEST PROCEDURE * *