20.2 In all measurements the P1 and P2 terminals are conventionally connected. Bear in
mind that thermal emfs may be present. The DO7e is particularly sensitive to these
on the 6m Ohm range.
20.3
For FS measurements, connect the C1 and C2 leads in the conventional manner (see
Fig
20.1).
20.4 For ZERO measurements, connect the C1 and C2 leads to the C2 terminal of the
standard (see Fig 20.2).
FULL SCALE CALIBRATION
P1, P2 LEADS - PURE COPPER, MULTI STRANDED
C1, C2 LEADS - 10A RATING (55/0.15MM)
Fig.
20.1
ZERO CALIBRATION
P1, P2 LEADS - PURE COPPER, MULTI STRANDED
C1, C2 LEADS - 10A RATING (55/0.15MM)
Fig.
20.2
* * END OF TEST PROCEDURE * *
Digital Microhmmeter Type DO7e
Operating Instructions
15
C1
P1
P2
C2
C1
P1
P2
C2
DO7
STANDARD RESISTOR
C1
P1
P2
C2
C1
P1
P2
C2
C1
P1
P2
C2
C1
P1
P2
C2
DO7
STANDARD RESISTOR
- AGL-5 (2 pages)
- Apollo 500 (21 pages)
- B433R (16 pages)
- Checkbox 17 Plus (4 pages)
- Apollo 600 (32 pages)
- B255 (40 pages)
- A203ES (10 pages)
- A252 (35 pages)
- DO4000 (22 pages)
- DO5000 Interface (41 pages)
- DO5000 Series (36 pages)
- DP6 (22 pages)
- DO7010 (44 pages)
- ARE 1557 (31 pages)
- Europa Pac Plus (53 pages)
- Supernova Elite (118 pages)
- HAL Scan (164 pages)
- G2000 (44 pages)
- IRT 1557 (21 pages)
- KD1E (22 pages)
- LLT (2 pages)
- PAC3760 (2 pages)
- PAC 1500Xi (12 pages)
- PAC 1000 (10 pages)
- PAT2000i (29 pages)
- PATGuard Elite Quick start (2 pages)
- PATGuard Elite SQL Quick start (2 pages)
- PATGuard Pro 2 (77 pages)
- PATguard Pro 2 Quick start (2 pages)
- PATGuard Time Manager (12 pages)
- PATGuard Work About Quick start (4 pages)
- PH3 LLT (6 pages)
- PATGuard Work About (53 pages)
- Solar Power Clamp (31 pages)
- PowerGuard Pro (29 pages)
- PowerPlus 1557 (62 pages)
- PowerPlus PC (44 pages)
- 100 PrimeTest (24 pages)
- PATGuard 3 (153 pages)
- 200 PrimeTest (34 pages)
- 250 PrimeTest (48 pages)
- 220 PrimeTest (34 pages)
- 50 PrimeTest (20 pages)
- RH9A Series (16 pages)