Rotary encoder alignment-using alignment tool, Installation instructions, Rotary encoders - using alignment tool - alignment – MicroE 1500V Mercury User Manual
Page 11
Installation Instructions
Rotary Encoders - Using Alignment Tool - Alignment
Page 9
5
Proper sensor alignment may require minor adjustments to the sensor position with
respect to the scale. This can be performed easily using the SmartPrecision Alignment
Tool as illustrated below.
The red, yellow, or green LED will light depending on sensor alignment. Slowly
move the sensor by allowing it to slide on the mounting surface until the green
LED, is illuminated. Optimal alignment will be displayed as a “Bright Green” LED.
IIM
MP
PO
OR
RT
TA
AN
NT
T:
Confirm that the green LED blinks when passing over
the index. If not, readjust the sensor in the Y direction and repeat the above
procedure. When alignment is completed, tighten the sensor mounting screws.
x
Y
Z
θ
z
To align the sensor, move it
in the Y or
θ
z directions.
6
SmartPrecision
Alignment Tool
Improper
Alignment LED
Red
Power/
Calibration
Power/
Calibration
Power/
Calibration
Improved
Alignment LED
Yellow
Proper
Alignment LED
Green
Optimal
Alignment LED
Bright Green
15 pin D
connector
to sensor
15 pin D
connector
to sensor
15 pin D
connector
to sensor
SmartPrecision
Alignment Tool
SSAT1000
SmartPrecision
Alignment Tool
SSAT1000
SmartPrecision
Alignment Tool
SSAT1000
Confirm green over the full range of
motion by sliding the scale past the
sensor. The “green” LED must
remain on over the entire range. If
not aligned over the entire range of
motion, loosen the sensor mounting
screws and repeat step 5.