Rotary encoder installation-using alignment tool – MicroE 1500V Mercury User Manual
Page 10
Installation Instructions
Rotary Encoders - Using Alignment Tool - Mounting
Attach your hub/scale
assembly to the rotary
device. Refer to the
interface drawing. The
reflective surface of the
scale must face the sensor.
1
Page 8
3
Be sure the source power is off
before connecting the SmartPrecision
Alignment Tool. (The M1500V-AC
Atmospheric Cable shown above is
optional. A customer-supplied cable
may be used.)
Connect the M1500V encoder to the
SmartPrecision Alignment Tool.
2
To M1500V encoder
To Power Supply
4
Connect the Alignment Tool to the
Computer Interface Adapter. Turn on
the power.
Install the sensor on your mounting surface
referencing the appropriate datum surface as
shown on the interface drawing. Use 2 washers
per mounting screw.
Benching pins may be used to locate the sensor
if the system mechanical tolerances are adequate.
See data sheet for alignment tolerances, or keep
mounting screws loose for sensor alignment if
benching pins are not used.